Product Highlights
The Yokogawa 701265 High Precision Temperature Input Module provides 2 channels of temperature (thermocouple) or voltage (switchable) measurement.
Specifications of the Yokogawa 701265 High Precision Temperature Input Module include:
- Function: Temperature (thermocouple) or voltage measurement (switchable)
- Effective measurement range: Voltage mesurement: twice of setting range
- Number of input channels: 2
- Input coupling:
- TC, DC, and GND
- TC: Temperature (thermocouple) measurement
- DC: Voltage measurement (DC coupling)
- Voltage measurement maximum sample rate: 100 kS/s
- Temperature measurement data update rate: 500 Hz
- Voltage:
- rms: 1.5 mV
- peak-to-peak: 10 mV
- Measurement range/accuracy:
- Voltage measurement:
- Voltage sensitivity: 1 mV to 100 V range (1-2-5 steps)
- Voltage accuracy: +/-(0.08% or range + 2u V)
- Frequency characteristics (-3 dB point when a sine wave of amplitude 60% of range is input):
- Voltage measurement: DC to 100 Hz
- Temperature measurement: DC to 100 Hz
- Maximum input voltage (at a frequency of 1 kHz or less):
- Temperature input: 42 V (DC + ACpeak)
- Voltage input: 42 V (DC + ACpeak)
- Maximum allowable common mode voltage (at a frequency of 1 kHz or less):
- Temperature input: 42 V (DC + ACpeak)(CAT l and CAT ll, 30 Vrms)
- Voltage input: 42 V (DC + ACpeak)(CAT l and CAT ll, 30 V)
- Vertical resolution:
- Voltage measurement during voltage input: 2400 LSB/div
- Temperature measurement when measuring temperature: 0.1C
- -3 dB point when AC coupled low frequency attenuation point: Voltage measurement 0.5 Hz or less
- Input connector: Binding post
- Input impedance: Approximately 1 MOhm
- Common mode rejection ratio:
- Voltage measurement: 80 dB (50/60 Hz) or more typical
- Temperature measurement: 120 dB or more (50/60 Hz, with 2-Hz filter ON, signal source resistance of 500 ohm or less) typical
- Residual noise level (input section shorted):
- Voltage measurement: +/-4 u V or +/-0.1% of range, whichever is greater (typical)
- A/D conversion resolution:
- Voltage measurement: 16 bits (24000 LSB/range)
- Temperature coefficient:
- Voltage measurement:
- Zero point: +/-(0.01% of range)/ C (typical)
- Gain: +/-(0.02% of range)/ C (typical)
- Bandwidth limit (digital filter): Select from OFF, 30 Hz, 8 Hz, and 2 Hz
- Input bias current: 20 nA or less