Response time: 120 ms (90 % energy) (adjustable via software)
Tariff fees or manufacturer imposed surcharges may be applicable.
Overview
The thermoMETER UC is an extremely powerful system for non-contact temperature measurement from -50 to +1000 C. It offers extensive functions and a modern design with standard M12 connections. The sensor is extremely small and can be used in environments with high temperatures up to 180 C. Due to its high performance and robustness, it is primarily used in machine building and machine design, in production processes and for quality assurance. Various interfaces such as analog and digital outputs or switching and alarm signals offer numerous possibilities for integration into processes, machines and systems.
Temperature measuring range: -50C to 900C
Optical resolution: 22:1
Spectral range: 8 to 14 µm
Emissivity: 0.95 (adjustable via software)
Response time: 120 ms (90 % energy) (adjustable via software)
Temperature resolution (NETD): 50 mK
System accuracy: +/-1.0 % or +/-1.0C (whichever is greater)
Ambient temperature: -20C to 180C (sensor), -20C to 85C (controller)
Supply: 5-36 VDC; 150 mA
Analog output: 0-5 V or 0-10 V, (0)4-20 mA (M12 5-pin connector)
Digital interfaces: RS485; 3.3 V-LVTTL for USB adapter; (Ethernet / EtherCAT /PROFINET / EtherNet/IP) optional; (M12 8-pin socket)
Switching outputs: 2 x relay (min/max; adjustable)
User interface: 2-line LCD display, 4 buttons; optional operation via sensorTOO software