Used Lecroy HDO6104-MS HD Mixed Signal Oscilloscope, 2.5Gs /s, 1GHz, 4+16 Channel, 50Mpts/Channel
HD Mixd.Sgnl.Oscilloscope,2.5Gs /s,1GHz,4+16Ch,50Mpts/Ch 12-Bi

Order #: HDO6104-MS-USED

Mfg #: HDO6104-MS-USED

Questions about availability or need help? Contact Us:

(800) 633-0561
$21,999.00

Used Lecroy HDO6104-MS HD Mixed Signal Oscilloscope, 2.5Gs /s, 1GHz, 4+16 Channel, 50Mpts/Channel

HD Mixd.Sgnl.Oscilloscope,2.5Gs /s,1GHz,4+16Ch,50Mpts/Ch 12-Bi

Order #: HDO6104-MS-USED

Mfg #: HDO6104-MS-USED

$21,999.00

Request a Used Instrument Quote

Have a Question? Speak to a Transcat Expert: 800.264.4059

Used instruments from TranscatTranscat Used Instrument Guarantee
Calibration Certificate Included 6 Month Warranty Unsurpassed Service & Support

Availability: Call for Availability
This product is available for shipping to the United States, Canada, and Puerto Rico only.
This product is available for shipping to the United States, Canada, and Puerto Rico only.
This product is available for shipping to the United States, Canada, and Puerto Rico only.

Product Highlights

  • 350 MHz, 500 MHz and 1 GHz bandwidths
  • Long Memory - up to 250 Mpts/Ch
  • Spectrum Analyzer Mode
  • WaveScan - Advanced Search and Find
  • History Mode - Waveform Playback
  • Power Analysis Software
  • JITKIT Clock-Data Jitter Analysis
  • Serial Data Trigger and Decode
  • PROTObus MAG Serial Data Debug Toolkit
  • LabNotebook Report Generation
  • Advanced Trigger Tools including TriggerScan
  • 16 Digital Channels with 1.25 GS/s Sample Rate
  • 12-bit ADC resolution, up to 15 bits with enhanced resolution

Combining Teledyne LeCroy's HD4096 high definition 12-bit technology, with long memory, a compact form factor, 12.1" touch screen display, powerful measurement and analysis tools, and mixed signal capability, the HDO6000 is the ideal oscilloscope for circuit validation, system debug and waveform analysis. The powerful feature set provides analytical tools and unique application packages to streamline the testing process. Tools such as WaveScan Search and Find and History Mode, combined with advanced triggering, identify and isolate problems while Spectrum Analyzer Mode provides analysis tools in the frequency domain.

HD4096 Technology

HD4096 high definition technology consists of high sample rate 12-bit ADCs, high signal-to-noise input amplifiers and a low-noise system architecture. This technology enables high definition oscilloscopes to capture and display signals of up to 1 GHz with high sample rate and 16 times more resolution than other oscilloscopes.

Long Memory

With up to 250 Mpts of memory the HDO6000 High Definition Oscilloscopes can capture large amounts of data with more precision than other oscilloscopes. The 2.5 GS/s, 250 Mpts architecture provides the ability to capture a fast transient or a long acquisition.

Large 12.1" Touch Screen

Navigating complicated user interfaces is a thing of the past thanks to the large touch screen display of the HDO6000. The user interface was designed for touch screens which makes navigating the HDO6000 extremely intuitive.

Comprehensive Analysis Tools

Advanced math and measurement parameters quantify andlog and digital waveforms while tracks, trends and histograms show how they change over time. Advanced triggering with TriggerScan and Measurement Trigger ensure even the most complicated signals are captured.

Win a Generac Generator when purchasing $750 or more from Transcat.Win a Generac Generator when purchasing $750 or more from Transcat.
Analog - Vertical  
Bandwidth @ 50 Ω (-3 dB) 1 GHz
Rise Time (10-90%, 50 Ω) 450 ps typical
Input Channels 4
Vertical Resolution 12-bits; up to 15-bits with enhanced resolution (ERES)
Sensitivity 50 Ω: 1 mV/div-1 V/div, fully variable 1 MΩ: 1 mV/div-10 V/div, fully variable
DC Gain Accuracy (Gain Component of DC Accuracy) ±(0.5%) F.S, offset at 0 V
Bandwidth Limiters 20 MHz, 200 MHz
Maximum Input Voltage 50 Ω: 5 Vrms, 1 MΩ: 400 V max (DC + Peak AC ≤ 10 Khz)
Input Coupling 50 Ω: DC, GND; 1 MΩ: AC, DC, GND;
Input Impedance 50 Ω ± 2.0%;1 MΩ ± 2.0% / 15 pF,
Offset Range 50 Ω: 1 mV - 4.95 mV: ±1.6 V, 5 mV - 9.9 mV: ±4 V, 10 mV - 19.8 mV: ±8 V, 20 mV - 1 V: ±10 V
1 MΩ: 1 mV - 4.95 mV: ±1.6 V, 5 mV - 9.9 mV: ±4 V, 10 mV - 19.8 mV: ±8 V, 20 mV - 100 mV: ±16 V
102 mV - 198 mV: ±80V, 200 mV - 1 V: ±160 V, 1.02 V -10 V: ±400 V
DC Vertical Offset Accuracy ±(1.0% of offset value + 0.5%FS + 0.02% of max offset + 1mV)
Analog - Acquisition  
Sample Rate (Single-shot) 2.5 GS/s
Sample Rate (Repetitive) 125 GS/s, user selectable for repetitive signals (20 ps/div to 10 ns/div)
Record Length Standard -STD: 50 Mpts/ch (all channels)
Option - L: 100 Mpts/ch (all channels)
Option -XL: 250 Mpts/ch (all channels)
Acquisition Modes Real-time, Roll, RIS (Random Interleaved Sampling), Sequence (Segmented Memory up to 30,000 segments, 60,000 segments -L Option, 65,000 -XL option) with 1us intersegment dead-time
Timebase Range 20 ps/div - 5 ks/div with standard memory (up to 10 ks/div with -L memory, 25 ks/div with -XL memory); RIS available at ≤ 10 ns/div; Roll Mode available at ≥ 100 ms/div and ≤ 5 MS/s
Timebase Accuracy ±2.5 ppm for 5 to 40C + 1.0ppm/year from calibration
Channel-Channel Deskew Range ±9 x time/div. setting, 100 ms max., each channel
External Timebase Reference (Input) 10 MHz ±25 ppm at 0 to 10 dBm into 50 Ω
External Timebase Reference (Output) 10 MHz 2.0 dBm ±1 dBm, sinewave synchronous to scope timebase
External Clock DC to 100 MHz; (50 Ω/1 MΩ), Ext. BNC input, Minimum rise time and amplitude requirements apply at low frequencies
Analog - Acquisition Processing  
Averaging Summed averaging to 1 million sweeps; continuous averaging to 1 million sweeps
Enhanced Resolution (ERES) From 12.5- to 15-bits vertical resolution
Envelope (Extrema) Envelope, floor, or roof for up to 1 million sweeps
Interpolation Linear (Default) or Sin x/X
Digital - Vertical and Acquisition (-MS Models Only)  
Input Channels 16 Digital Channels
Threshold Groupings Pod 2: D15 - D8, Pod 1: D7 - D0
Threshold Selections TTL, ECL, CMOS (2.5 V, 3.3 V, 5 V), PECL, LVDS or User Defined
Maximum Input Voltage ±30V Peak
Threshold Accuracy ±(3% of threshold setting + 100mV)
Input Dynamic Range ± 20V
Minimum Input Voltage Swing 400mV
Input Impedance (Flying Leads) 100 kΩ / 5 pF
Maximum Input Frequency 250 MHz
Sample Rate 1.25 GS/s
Record Length Standard -STD: 50 MS - 16 Channels
Optional -L: 100 MS - 16 Channels
Optional -XL: 125 MS - 16 Channels
Minimum Detectable Pulse Width 2 ns
Channel-to-Channel Skew 350 ps
User Defined Threshold Range ±10 V in 20 mV steps
User Defined Hysteresis Range 100 mV to 1.4 V in 100 mV steps
Triggering System  
Modes Auto, Normal, Single, Stop
Sources Any input channel, External, Ext/10, or line; slope and level unique to each source (except for line trigger)
Coupling DC, AC, HFREJ, LFREJ
Pre-trigger Delay 0-100% of memory size
Post-trigger Delay 0-10,000 Divisions in real time mode, limited at slower time/div settings or in roll mode
Hold-off From 2 ns up to 20 s or from 1 to 99,999,999 events
Internal Trigger Level Range ±4.1 div from center (typical)
External Trigger Input Range Ext: ±400 mV, Ext/10: ±4 V
Maximum Trigger Rate 1M Triggers/sec (in Sequence Mode, up to 4 channels)
Trigger Sensitivity with Edge Trigger (Ch 1-4) 0.9 division: 10 MHz
1.0 divisions: 200 MHz
1.5 divisions: 500 MHz
2.0 divisions: 1 GHz
External Trigger Sensitivity, (Edge Trigger) 0.9 division: 10 MHz
1.0 divisions: 200 MHz
1.5 divisions: 500 MHz
2.0 divisions: 1 GHz
Max. Trigger Frequency, (C1-C4, Aux In, Smart Trigger) 1 GHz
Trigger and Interpolator Jitter ≤ 3.5 ps rms (typical)
<0.1 ps rms (typical, software assisted)
Trigger Types  
Edge Triggers when signal meets slope (positive, negative, or either) and level condition
Width (Signal or Pattern) Triggers on positive or negative glitches with selectable widths selectable as low as 200 ps (depending on oscilloscope bandwidth); Maximum Width: 20 s
Pattern Logic combination (AND, NAND, OR, NOR) of 5 inputs (4 channels and external trigger input). Each source can be high, low, or don't care. The High and Low level can be selected independently. Triggers at start or end of the pattern
Measurement Trigger Trigger on Measurement with qualified limits.
TV-Composite Video Triggers NTSC or PAL with selectable line and field;
HDTV (720p, 1080i, 1080p) with selectable frame rate (50 or 60 Hz) and Line; or
CUSTOM with selectable Fields (1-8), Lines (up to 2000), Frame Rates (25, 30, 50, or 60 Hz),
Interlacing (1:1, 2:1, 4:1, 8:1), or Synch Pulse Slope (Positive or Negative)
Smart Triggers  
Window Triggers when signal exits a window defined by adjustable thresholds
Interval (Signal or Pattern) Triggers on intervals selectable between 1 ns and 20 s
Glitch Triggers on positive or negative glitches with widths selectable as low as 200 ps (depending on oscilloscope bandwidth) to 20 s, or on intermittent faults
Dropout Triggers if signal drops out for longer than selected time between 1 ns and 20 s
Runt Trigger on positive or negative runts defined by two voltage limits and two time limits. Select between 1 ns and 20 ns
Slew Rate Trigger on edge rates. Select limits for dV, dt, and slope. Select edge limits between 1 ns and 20 ns
Multi-Stage Triggers  
Cascade (Sequence) Triggering  
Capability Arm on “A" event, then Trigger on “B" event. Or Arm on “A" event, then Qualify on “B" event, and Trigger on “C" event. Or Arm on “A" event, then Qualify on “B" then “C" event, and Trigger on “D" event
Types A, B, C, or D event: Edge, Glitch, Width, Window, Dropout, Interval, Runt, Slew Rate, Pattern (analog), or Measurement.
Holdoff Holdoff between A and B, B and C, C or D, or any is selectable by time or number of events
Qualified First In Sequence acquisition mode, triggers repeatably on event B only if a defined pattern, state, or edge (event A) is satisfied in the first segment of the acquisition. Holdoff between sources is selectable by time or events
Qualified Triggers on any input source only if a defined state or edge occurred on another input source. Delay between sources is selectable by time or events
TriggerScan A Trigger Trainer analyzes the waveforms, identifies normal behavior, and then sets up a large set of rare event smart trigger setups that target abnormal behavior. The trainer ‘learns' trigger setups based on slew rates, periods, amplitudes outside of a range and then applies them sequentially.
Triggers with Exclusion Technology Glitch, Width, Interval, Runt, Slew Rate - Trigger on intermittent faults by specifying the expected behavior and triggering when that condition is not met
Measurement Tools  
Measurement Functionality Display any 8 parameters together with statistics, including their average, high, low, and standard deviations. Histicons provide a fast, dynamic view of parameters and wave shape characteristics. Parameter Math allows addition, subtraction, multiplication, or division of two different parameters. Parameter gates define the location on the source waveform. Each occurrence of each parameter is measured and added to the statistics table.
Measurement Parameters Amplitude, Area, Base (Low), Cycles, Data, Delay, Delta Delay, Duty Cycle, Duration, Fall time (90-10%, 80-20%, @ level), Frequency, First, Last, Level @ x, Maximum, Mean, Median, Minimum, Narrow band phase, Narrow band power, Number of points, + Overshoot, - Overshoot, Peak-to-peak, Period, Risetime (10-90%, 20-80%, @ level), RMS, Std. deviation, Top, Width, Median, Phase, Time @ minimum (min.), Time @ maximum (max.), Delta time @ level, Delta time @ level from trigger, X @ max., X @ min., Cycle-Cycle Jitter, N-Cycle, N-Cycle with start selection, Frequency @ level, Period @ level, Half Period, Width @ level, Time Interval Error @ level, Setup, Hold, Skew, Duty Cycle @ level, Duty Cycle Error, Edge @ lv (counts edges)
Math Tools  
Math Functionality Display up to 8 math function traces (F1-F8). The easy-to-use graphical interface simplifies setup of up to two operations on each function trace, and function traces can be chained together to perform math-on-math.
Math Operators Absolute value, Average (summed), Average (continuous), Correlation (two waveforms), Cubic interpolation, Derivative, Deskew (resample), Difference (-), Enhanced resolution (to 15 bits vertical), Envelope, Exp (base e), Exp (base 10), FFT (power spectrum, magnitude, phase, power density, real, imaginary, magnitude squared, up to 128 Mpts and rectangular, VonHann, Hamming, FlatTop and Blackman Harris windows), Floor, Integral, Interpolate (cubic, quadratic, sinx/x), Invert (negate), Log (base e), Log (base 10), Product (x), Reciprocal, Rescale (with units), Roof, (SINx)/x, Sparse, Square, Square root, Sum (+), Zoom (identity). 2 dual operator math functions may be defined at a time.
Measurement and Math Integration  
  Histograms expanded with 19 histogram parameters and up to 2 billion events. Trend (datalog) of up to 1 million events. Track graphs of all parameters. Persistence histogram, persistence trace (mean, range, sigma)
Pass/Fail Testing  
Test Types Parameter limit testing, mask testing. Pass/Fail Actions include: Save, Stop, Alarm, Pulse, Hardcopy, LabNotebook
Probes  
Standard Probes PP018 (5 mm) (Qty. 4)
Probing System BNC and Teledyne LeCroy ProBus for Active voltage, current and differential probes
Display System  
Display Size Color 12.1" widescreen flat panel TFT-Active Matrix with high resolution touch screen
Display Resolution WXGA; 1280 x 800 pixels
Number of Traces Display a maximum of 16 traces. Simultaneously display channel, zoom, memory and math traces
Grid Styles Auto, Single, Dual, Quad, Octal, X-Y, Single+X-Y, Dual+X-Y, Tandem, Quattro, Twelve, Sixteen
Waveform Representation Sample dots joined, or sample dots only
Connectivity  
Ethernet (2) 10/100/1000Base-T Ethernet interface (RJ-45 connector)
USB Host Ports (6) USB Ports Total - (2) Front USB Ports
USB Device Port (1) USBTMC Port
GPIB Port (Optional) Supports IEEE - 488.2 (External)
External Monitor Port Standard 15-pin D-Type SVGA-compatible DB-15 connector, DVI connector and HDMI connector
Remote Control Via Windows Automation, or via Teledyne LeCroy Remote Command Set
Processor/CPU  
Type Intel Core i5, 2.5 GHz (or better)
Processor Memory 8 GB standard
Operating System Windows® Embedded Standard 7 Professional, 64-bit
Power Requirements  
Voltage 100-240 VAC ±10% at 45-66 Hz; 100-120 VAC ±10% at 380-420 Hz; Automatic AC Voltage Selection; Installation Category: 300 V CAT II
Power Consumption (Nominal) 200 W / 200 VA
Max Power Consumption 350 W / 350 VA (with all PC peripherals and active probes connected to 4 channels)
Environmental  
Temperature Operating: 5 °C to 40 °C; Non-Operating: -20 °C to 60 °C
Humidity Operating: 5% to 90% relative humidity (non-condensing) up to +31 °C, Upper limit derates to 50% relative humidity (non-condensing) at +40 °C; Non-Operating: 5% to 95% relative humidity (non-condensing) as tested per MIL-PRF-28800F
Altitude Operating: 3,048 m (10,000 ft) max at ≤ 30C; Non-Operating: Up to 12,192 meters (40,000 ft)
Random Vibration Operating : 0.31 grms 5 Hz to 500 Hz, 15 minutes in each of three orthogonal axes; Non-Operating: 2.4 grms 5 Hz to 500 Hz, 15 minutes in each of three orthogonal axes
Functional Shock 30 gpeak, half sine, 11 ms pulse, 3 shocks (positive and negative) in each of three orthogonal axes, 18 shocks total
Physical  
Dimensions (HWD) 11.48"H x 15.72"W x 5.17"D (291.7 mm x 399.4 mm x 131.31 mm)
Weight 5.86 kg (12.9 lbs)
Certifications  
CE Certification Low Voltage Directive 2006/95/EC
EN 61010-1:2010, EN 61010-2-030:2010
EMC Directive 2004/108/EC
EN 61326-1:2006, EN61326-2-1:2006
UL and cUL Listing UL 61010-1 (3rd Edition), UL 61010-2-030 (1st Edition); CAN/CSA C22.2 No.61010-1-12
Warranty and Service  
  3-year warranty; calibration recommended annually. Optional service programs include extended warranty, upgrades, and calibration services