Order #: 5790B-03-RENT2
Mfg #: 5790B-03-RENT2
Fluke 5790B-03 FOR RENT
Order #: 5790B-03-RENT2
Mfg #: 5790B-03-RENT2

Product Highlights
the 5790B AC Measurement Standard is a multi-purpose ac measurement and transfer standard designed for the most demanding calibration applications. It combines the accuracy you would expect from a thermal transfer standard with the ease of use of a digital multimeter. Absolute ac voltage measurement uncertainties are as low as ± 24 ppm (one year, 23 °C ± 5 °C). The 5790B is designed to meet the complete ac voltage, ac current and wideband verification requirements of the fluke Calibration 5735A, 5720A, and 5700A Multifunction Calibrators; 5522A, 5502A, 5520A, and 5500A Multi-Product Calibrators; plus other calibrators, amplifiers like the 52120, 5725A and 5205A/5215A, and transfer standards and ac voltmeters.
The 5790B covers an alternating voltage range of 600 µV to 1000 V, and a frequency range of 10 Hz to 1 MHz. Wideband voltage options extend frequency range to 30 or 50 MHz.
The 5790B in combination with the fluke A40B Current Shunts now allow you to make direct, absolute or relative current measurements without the need to perform any calculations.
The 5790B may be used alone as a measurement device or as a transfer standard when compared to an external dc source. In either case the normally tedious switching and calculations are performed automatically by the 5790B, and the resulting ac/dc difference is displayed directly on the easy-to-read 6.5-inch full-color capacitive touch screen display.
Summary Specifications | |||||
AC and dc voltage |
|
||||
Inputs |
Input 1: 600 µV to 1000 V (5% over-range) Input 2: 600 µV to 1000 V (5% over-range) Aux input: 250 mV to 500 mV WB input: <7 V rms wideband |
||||
Warm-up time | 30 minutes | ||||
Relative humidity |
Operating: 45% to 50 °C 75% to 45 °C 95% to 30 °C Storage: <95% non-condensing |
||||
Altitude |
Operating: 3,050 meters (10,000 feet) Non-operating: 12,200 meters (40,000 feet) |
||||
Temperature |
Operating: 0 °C to °50 C Calibration: 15 °C to 35 °C Storage: -40 °C to 70 °C |
||||
Electromagnetic Compatibility (EMC) | |||||
International | IEC 61326-1: Controlled Electromagnetic Environment - CISPR 11: Group 1, Class A | ||||
Korea (KCC) | Class A Equipment (Industrial Broadcasting & Communication Equipment) | ||||
USA (fCC) | 47 CfR 15 subpart B. This product is considered an exempt device per clause 15.103. | ||||
Surge | ANSI C62.41-1980, Category A | ||||
Reliability | MIL-T-2880D, paragraph 3.13.3 | ||||
General Specifications | |||||
Dimensions |
Height: 17.8 cm (7 in) standard rackmount + 1.5 cm (0.6 in) Width: 43.2 cm (17 in) Depth: 62 cm (24.8 in) |
||||
Maximum power requirements |
5790B: 100 VA |
||||
Weight |
5790B: 24 kg (53 lb) With wideband: 24.5 kg (54 lb) |
||||
Line power |
50 Hz or 60 Hz; 100 V - 120 V, 220 V - 240 V 100V, 110 V, 115 V, 120 V, 200 V, 220 V, 230 V, 240 V |
||||
Safety | Complies with UL1244 and IEC 348-1976 and IEC 1010 and CSA C22.2 No. 231 and ANSI/ISA S82 | ||||
Remote interfaces | RS-232, IEEE-488 | ||||
Confidence level | 99% unless otherwise specified | ||||
5790B/Af | |||||
The 5790B/Af has enhanced absolute accuracy to ± 0.23 % of voltage (1 year after factory calibration, 23 °C ± 3 °C, at 95 % confidence level) at 50 MHz and 223.61 mV, using a serialized one meter cable, characterized at that point for Rf power meter 0 dBm reference output measurement. In the Rf power measurement mode, the accuracy is specified to the end of the serialized cable with a curve fit for the other ranges and frequencies that is a first order correction for cable loss. |
Special Member Pricing
Buy with Calibration
Rental Instruments
Access to 100,000+ products
1,000's of instruments in stock
Product selection expertise
Application assistance
Purchase with a PO
- North Atlantic 8810AR FOR RENTAngle Position Indicator with On-Board Ref.,ProgrammableCall for price
-
-
-
-
-
- Fluke Calibration 9640A FOR RENT4GHz Ultra Low Phase Noise RF Ref.SourceOutput, 50ohmCall for price
-
- North Atlantic NORTH-ATLANTIC 8810A for RentAngle Position Indicator w/On-board Ref., ProgrammbleCall for price
-
-
-
- Fluke Calibration 5720A-03 for Rent
High Performance Multifunction Calibrator with Wideband Option
Call for price
How to Rent from Transcat

1. Select Your Rental Product
- Need help selecting a unit? Try using our Product Finder or search our available units
- Can’t find what you are looking for? Is the rental unit you need, not available? Give us a call at 800-828-1470 or our chat feature to speak with one of our dedicated Rental Experts

2. Contact Our Team of Rental Experts
- Contact us here or click below to submit your request for a rental quote and a member of our Rental Team will review the request and will reach out to you very soon

3. Select Your Rental Period & Complete Your Order
- Our weekly period lasts 7 calendar days, while our monthly rental period lasts 28 calendar days
- Provide your form of payment, sign the rental agreement and any other required documents, and we will ship you the unit