Overview

  • Voltage DC Range: 0 to ±1020V
  • Best 1 Year Specification: ±50 ppm of setting
  • Current DC Range: 0 to ±11A
  • Best 1 Year Specification: ±0.01% of setting
  • Resistance Range: 0 to 3.29.999 MΩ
  • Best 1 Year Specification: ±0.009% of setting
  • Voltage AC Range: 1 mV to 1020V, 10 Hz to 500 kHz, Sine
  • Best 1 Year Specification: ±0.03% of setting
  • Current AC Range: 29 m A to 11A, 10 Hz to 10 kHz, Sine
  • Best 1 Year Specification: ±0.06% of setting
  • Capacitance Range: 0.33 nF to 1.1 mF
  • Best 1 Year Specification: ±0.25% of setting
  • Thermocouple Source and Measure Range: -250°C to +2316°C (11 types)
  • Best 1 Year Specification: ± 0.14°C of setting
  • RTD Source Range: -200°C to +630°C(8 types)
  • Best 1 Year Specification: ±0.03°C of setting
  • DC Power Range: 109 mW to 11 kW
  • Best 1 Year Specification: ± 0.08% of setting
  • AC Power Range: 109 mW to 11 kW, 45-65 Hz, PF=1
  • Best 1 Year Specification: ± 0.15% of setting
  • Phase Range: 0 to ± 179.99°
  • Best 1 Year Specification: ±0.15°
  • Frequency Range: 0.01 Hz to 2.0 MHz
  • Best 1 Year Specification: ±25 ppm of setting
  • Warm up Time: Twice the time since last warmed up, to a maximum of 30 minutes
  • Settling Time: Less than 5 seconds for all functions and ranges except as noted
  • Standard Interfaces: IEEE-488 (GPIB), RS-232, 5725A (5500A only)
  • Temperature Performance Operating: 0°C to 50°C
  • Calibration: (tcal): 15°C to 35°C
  • Storage: -20°C to 70°C
  • Temperature Coefficient: Temperature coefficient for temperatures outside tcal ± 5°C is 10% of the 90-day specification (or 1-year, as applicable) per °C
  • Relative Humidity Operating: < 80% to 30°C, < 70% to 40°C, < 40% to 50°C
  • Storage: < 95%, non-condensing
  • Altitude Operating: 3,050m (10,000 ft) maximum
  • Non-operating: 12,200m (40,000 ft) maximum
  • Safety: Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994; CAN/CSA-C22.2 No. 1010.1-92
  • Analog Low Isolation: 20V
  • EMC: Designed to comply with FCC Rules Part 15
  • Line Power: Line Voltage (selectable): 100V, 120V, 220V, 240V
  • Line Frequency: 47 to 63 Hz
  • Line Voltage Variation: ± 10% about line voltage setting
  • Power Consumption: 300 VA
  • Calibration Documentation: 17025 accredited report of calibration included
  • Dimensions Height: 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on bottom of unit
  • Width: 43.2 cm (17 in), standard rack width
  • Depth: 47.3 cm (18.6 in) overall
  • Weight: (without options) 20 kg (44 lbs.)
  • Absolute Uncertainty Definition: 5500A and 5520A uncertainty specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. You do not need to add anything to determine the total uncertainty of your calibrator for the temperature range indicated.
  • Specification Confidence Interval: > 99%

Technical Resources