ADTECH SCX202-H30 DC: I/V/mV Isolated XMTR with T DIN "G" Rail 20" long
DC: I/V/mV Isolated XMTR with T DIN "G" Rail 20" long

Order #: SCX202-H30

Mfg #: SCX202-H30

Questions about availability or need help? Contact Us:

(800) 633-0561
$374.00

ADTECH SCX202-H30 DC: I/V/mV Isolated XMTR with T DIN "G" Rail 20" long

DC: I/V/mV Isolated XMTR with T DIN "G" Rail 20" long

Order #: SCX202-H30

Mfg #: SCX202-H30

$374.00
Availability: Ships in 3-4 weeks

Choose Your Options:

Add Calibration

You have chosen to add the following Calibration Services to Your Order

Make sure your instrument is calibrated and ready to use the minute you open the box by adding our expert calibration service. Why Calibrate? For a complete descriptions of Transcat Service Levels, click here.

Already have this this item and need calibration? Request a calibration quote

Input Signals

  • 4-20 mA DC (z in 10 ohms)
  • 0-20 or ± 20 mA DC (z in 10 ohms)
  • 0-10 or ±10 mA DC (z in 20 ohms)
  • 0-1 or ±1 mA DC (z in 200 ohms)
  • 1-5 VDC (z in 1 megohm)
  • 0-5 or ±VDC (z in 1 megohm)
  • 0-10 or ±10 VDC (z in 1 megohm)
  • Any unipolar or bipolar voltage from 100 MV to 200 VDC. (Option I 14)
  • Zero Suppression: ±10%
  • Span Adjustment: ±10%

Performance

  • Calibrated Accuracy: ±0.1%
  • Independent Linearity: ±0.025% maximum, ±0.01% typical
  • Repeatability: ±0.005% max., ±0.002% typ.
  • Zero TC: ±0.007% of span max./°C
  • Span TC: ±0.008% of span max. / °C
  • Load Effect: ±0.005% zero to full load
  • Output Ripple: 10 mV P/P maximum
  • Response Time: 110 milliseconds ( 10 to 90% step response)
  • Bandwidth: (-3 db): 3.2 Hz
  • Temperature Range: -25° to 185°F (-31° to 85°C) operating; -40° to 200°F (-40° to 93°C) storage
  • Power Supply Effect: ±0.005% over operating Range
  • Isolation: Input/output/case: 1000 VDC, 600 VAC

Power:

  • 8 to 42 VDC: Standard

Ordering Options:

  • H 15 D, H 25 - H 30 Mounting
  • LPI 40 D Loop powered indicator
  • I 14 Voltage / current inputs