PRT, SPRT and System Calibrations
Transcat also offers Probe-only and System Calibrations (w/ probe and readout together) for reference grade PRT's (Platinum Resistance Thermometers), Precision Thermistors, and SPRT's (Standard Platinum Resistance Thermometers). All calibrations are traceable through extremely stable and uniform Temperature Baths/Blocks, the Triple Point of Water, Mini Fixed Point Cells, including Indium (156.5985 °C), Tin (231.928 °C), Zinc (419.527 °C), aluminum (660.323 °C), SPRT's and a DC Resistance Bridge, which offers a temperature range of -196°C to 660.323°C and a best-case uncertainty of 0.6 millikelvin in our Dayton lab and 0.5 millikelvin in the Philadelphia lab.
Upon realization of the ITS-90 scale, as outlined by NIST Technical Note 1265 (Guidelines for Realizing the International Temperature Scale of 1990), and by utilizing the above techniques, the customer is provided with ITS-90 Coefficients pertaining to the temperature range calibrated. Temperature vs. resistance tables will be provided, depending on the level of service requested and/or provided.
Hart 5615-X Series Secondary Industrial Probe |
Hart 5626/5628 Series SPRT |
Hart 5610-X Thermistor Probe |
Kaye Instruments IRTD |
Ametek STS Series PRT |
Thermoprobe Tl1-A Thermometer |
Fluke 1551/1552ex Stik Thermometer |
Hart 5627a-X Series Secondary Industrial Probe |
Hart 1521/1522 Handheld Thermometer |
Hart 1523/1524 Handheld Thermometer |
Hart 1502A Tweener Readout |
Hart 1529-X Chub E-4 Readout |
Ametek DTI Series Handheld Thermometer |
|
Hart 914x Series Metrology Well |
Keithley 2182 NanoVolt Meter |
Ametek RTC/ATC Series Dry Block |
Kaye K140 Ice Point Reference |