Overview :

Product Highlights

Hioki LCR Meters and Impedance Analyzers range from 1mHz to 3GHz devices to suit a wide range of applications in the testing of electronic components. The IM3570 is ideal for use in applications requiring low-ESR measurement on the order of several milliohms, for example testing of functional polymer capacitors, due to its superior low-impedance repeatability.

  • LCR, DCR, sweep, and continuous measurement with high-speed testing in a single unit
  • Maximum test speeds of 1.5ms (1 kHz) and 0.5ms (100kHz) in LCR mode
  • High basic accuracy of Z parameter: ±0.08%
  • Ideal for testing the resonance characteristics of piezoelectric elements, C-D and low ESR measurement of functional polymer capacitors, DCR and L-Q measurement of inductors (coils and transformers)
  • Perform frequency sweeps, level sweeps, and time interval measurements in analyzer mode
  • Support for high-speed switching of measurement conditions

Specifications :

Measurement modesLCR mode, Analyzer mode (Sweeps with measurement frequency and measurement level), Continuous measurement mode
Measurement parametersZ, Y, θ, Rs (ESR), Rp, Rdc (DC resistance), X, G, B, Cs, Cp, Ls, Lp,
D (tanδ), Q
Measurement range100 mΩ to 100 MΩ, 12 ranges (All parameters are determined according to Z)
Display rangeZ, Y, Rs, Rp, Rdc, X, G, B, Ls, Lp, Cs, Cp :
±(0.000000 [unit] to 9.999999G [unit], Absolute value display for Z and Y only
θ : ±(0.000° to 180.000°), D : ±(0.000000 to 9.999999)
Q : ±(0.00 to 99999.99), Δ % : ±(0.0000% to 999.9999%)
Basic accuracyZ ±0.08% rdg. θ: ±0.05°
Measurement frequency4 Hz to 5 MHz (5 digits setting resolution, minimum resolution 10 mHz)
Measurement signal levelNormal mode:
V mode/CV mode: 5 mV to 5 Vrms (up to 1 MHz)
10 mV to 1 Vrms (1.0001 MHz to 5 MHz), 1 mVrms steps
CC mode: 10 μA to 50 mArms (up to 1 MHz)
10 μA to 10 mArms (1.0001 MHz to 5 MHz), 10 μArms steps
Low impedance high accuracy mode:
V mode/CV mode: 5 mV to 1 Vrms (up to 100 kHz), 1 mVrms steps
CC mode:10 μA to 100 mArms (100 mΩ and 1Ω ranges of up to 100 kHz), 10 μArms steps
Output impedanceNormal mode: 100 Ω, Low impedance high accuracy mode: 10 Ω
Display5.7-inch color TFT, display can be set to ON/OFF
Measurement time0.5 ms (100 kHz, FAST, display OFF, representative value)
FunctionsDC bias measurement, Comparator, BIN measurement (classification), Panel loading/saving, Memory function
InterfacesEXT I/O (handler), RS-232C, GP-IB, USB communication, USB memory, LAN
Power supply90 to 264 V AC, 50/60 Hz, 150 VA max.
Dimensions and mass330 mm (12.99 in) W × 119 mm (4.69 in) H × 307 mm (12.09 in) D, 5.8 kg (204.6 oz)
AccessoriesPower cord × 1, Instruction manual × 1, PC communication instruction manual (CD-R) × 1